CASCADA III.I SYSTEM
Output Flow Rates
|
|||
System |
RO |
Type III |
Type I |
Cascada III.I 5 |
5 L/h |
≥ 2 L/min* |
Up to 2 L/min |
Cascada III.I 10 |
10 L/h |
≥ 2 L/min* |
Up to 2 L/min |
Cascada III.I 20 |
20 L/h |
≥ 2 L/min* |
Up to 2 L/min |
Cascada III.I 30 |
30L/h |
≥ 2 L/min* |
Up to 2 L/min |
Output Water Quality
Water Type lll
Resistivity (MΩ • cm @ 25ºC) : Ion Rejection rate >99%
TOC(ppb) : Organic Rejection Rate > 99%
Bacteria (cfu/mL) : <1*
Particles (>0.2 μm) : <1/mL*
Endotoxin (Eu/mL) : –
RNases (ng/mL) : –
DNases (pg/μL) : –
Water Type l
Resistivity (MΩ • cm @ 25ºC) : 18.2
TOC(ppb) : < 5 ***
Bacteria (cfu/mL) : <1*
Particles (>0.2 μm) : <1/mL*
Endotoxin (Eu/mL) : < 0.001 **
RNases (ng/mL) : < 0.01 **
DNases (pg/μL) : < 4 **
* With 0.2 μm bacteria removal final filter
** With positive charged Nylon KA1 endotoxin removal final filter
*** With feed TOC less than 2 ppm
Feed Water Requirements
Conductivity : <1500 μs/cm @ 25 ºC
Pressure : 0.5 ~ 6 bar
Temperature : 5 ~ 35 ºC
Free Chlorine : < 3 ppm
Silt Density Index : <12
pH : 4 ~ 10
Dimension (mm)
System : H 575 x W 366 x D 492
Reservoir : H1200/900/600 x W 390 x D 384
Pre-Treatment : H 463 x W 220 x D 380
Dispensiong Station : H 845 x W 280 x D 300
Net Weight (kg)
System : 23
Reservoir : 5(35L) ; 7(70L) ; 9 (105L)
Pre-Treatment : 7
Dispensiong Station : 6
Electrical Requirements
Input Voltage : 100-240 V 50-60 Hz
Power : 200 VA Main Unit
75 VA Pretreatment