ELGA PURELAB Option Q 15 (With Boost Pump)

The PURELAB Option Q 15 produces ultra pure water without the need for pre-purified feed. It can produce up to 15 l/hr of 18.2MΩ-cm water direct from a potable water supply

AAS (Atomic absorption spectrophotometry), Bacterial cell culture, Buffer and media preparation, Clinical biochemistry, Electrochemistry, Electrophoresis, Electrophysiology, ELISA (Enzyme linked immunoabsorbant assay), Endotoxin analysis, Flame-AAS, GC-MS, General chemistry, GF-AAS (Graphite Furnace – Atomic Absorption Spectrophotometry), Histology, HPLC, Hydroponics, IC (Ion chromatography), ICP-AES (Inductively Coupled Plasma – Atomic Emission Spectrometry), ICP-MS (Inductively Coupled Plasma – Mass Spectrometry), Immunocytochemistry, Mammalian and bacterial cell culture, Media preparation, Microbiological analysis, Molecular biology, Monoclonal antibody research, Plant growth cabinets, Plant tissue culture, Qualitative analyses, RIA (Radioimmunoassay), Sample dilution and reagent preparation, Solid phase extraction, Spectrophotometry, Stability chambers, TOC analysis, Trace metal detection, Ultra sonic bath, Ultra trace analysis, Water analysis

Make-Up Rate 15 l/hr
Delivery Flow Rate up to 1 l/min
Daily Usage (8hr day) < 100 l/day
Inorganics 18.2 MΩ-cm
Organics (TOC) <3 ppb
Bacteria <0.1 CFU/ml (<1 CFU/10ml)
Endotoxins <0.001 EU/ml
DNase <20 pg/ml
RNase 0.002 ng/ml
Footnotes Bacteria – Specification subject to use with a POU (point of use) Biofilter. If Biofilter not used, Bacteria level is <1 CFU/ml
TOC – Specification dependant on feedwater
Endotoxin – Specification subject to use with a Biofilter
DNase/RNase – Specification subject to use with a Biofilter

 

Feedwater Nature Potable supply (Tap Water)
Conductivity <1400 µS/cm
Fouling Index <10
Total Chlorine 0.5 ppm
Silica 30 ppm
Temperature 1-35 °C
Flow Rate <85 l/hr
Drain (gravity fall with air gap) 70 l/hr
Min. Pressure Flooded suction with internal boost pump
Max. Pressure 2 bar (30 psi) with internal boost pump