Name: N2014 Starter Kit (Thermocouple)

SKU: N2014Starter Kit
Description:
Diligence EV Thermocouple Data Logger with LCD Display

Diligence EV data loggers are lightweight and small enough to be used almost anywhere. Ideal for many applications including the food, pharmaceutical and chemical industries to monitor sensitive materials in transit or storage. The loggers are also ideal for building services, scientific experiments, production processes and heating, ventilating and air conditioning systems.

Requires dedicated software and a computer interface to download data to a PC. Two versions of software are available, EVSW for general use and ideal for HACCP and other quality systems, EVSWPRO which aids compliance with 21CFR Part 11.

For single channel operation select a probe with a Lumberg connector – L suffix.

For connection up to three channels use Adaptor N2000ADP/K for Type K or N2000ADP/T for Type T and select probes with a sub-miniature connector – suffix M.

  • Features four channels for up to 3 external thermocouple probes or one external thermistor probe or one external thermocouple probe with thermistor at the cold junction
  • Wide measurement ranges from -200°C to thermocouple limits, 400°C type T and 1372°C type K
  • LCD Display for temperature readings and alarm indication
  • Tough, moulded case, dust and waterproof to IP67 standards
  • Large memory capacity of up to 16000 readings
  • Ability to log over multiple periods
  • Single, multi-function button
  • LED indicators to show active logging and alarm conditions, plus audible alarm warning
  • Temperature system accuracy, suitable for food industry applications
  • Long battery life
  • Two year warrantyKit includes:
  • 1 x N2014 Data Logger
  • 1 x N2000CRU Computer Interface
  • 1 x EVSW EV SoftwareEverything you need to start logging!

Detail:

Sensor Type Thermocouple type T
Thermocouple type K
Thermistor

Measurement Range
Type K -200°C to +1372°C/-328°F to +2502°F
Type T -200°C to +400°C/-328°F to +752°F
Thermistor -40°C to +70°C/-40°F to +158°F

Scales °C and °F
Resolution (Thermocouple) 0.2°C up to and including +300°C
0.5°C above +300°C

Temperature accuracy
+20°C ambient ±0.5°C +0.3% of reading (Logger Only)
-20°C to +40°C ambient ±0.5°C +0.5% of reading (Logger Only)
-40°C to +70°C ambient ±1.0°C +1.0% of reading (Logger Only)
-20°C to +40°C ambient ±0.3°C Thermistor (System Accuracy with Thermistor)
-40°C to +70°C ambient ±0.5°C Thermistor (System Accuracy with Thermistor)

Memory 32K
1 channel 16000 samples
2 channels 8000 samples
3 channels 5300 samples
4 channels 4000 samples

Communications Via infra red interface
Download Time 3 minutes for 10,000 readings (typical)
Logging Frequency Programmable between 2 seconds and 99 hours
Internal Sensor Response time T90 = 15 to 30 minutes in ambient air
Ambient Storage -40°C to +70°C/-40°F to +158°F

Logging Start Programmable date and time or manual start from instrument multi-function button
Software Windows™ based software for Windows XP or Vista

Battery Type 1 x AA size 3.6V replaceable lithium battery
Battery Life Up to 5 years

Case Material ABS
Environmental Protection IP67 BS EN 60529 IEC 529

LED Indication
Red Alarm
Green Logger Active

Dimensions
Length 88mm/3.5ins (including fixing lug)
Width 80mm/3.1ins
Depth 35mm/1.4ins

EVSW Software and New USB Cradle Drivers can be downloaded using these links.

Diligence EV Thermocouple Data Logger with LCD Display Accessories

N2000CRU USB Computer Interface
A17476 Lithium Battery
N2000ADP/K K Type Adapter for N2014
N2000ADP/T T Type Adapter for N2014