Name: N2014 Starter Kit (Thermocouple)

Diligence EV Thermocouple Data Logger with LCD Display
Diligence EV data loggers are lightweight and small enough to be used almost anywhere. Ideal for many applications including the food, pharmaceutical and chemical industries to monitor sensitive materials in transit or storage. The loggers are also ideal for building services, scientific experiments, production processes and heating, ventilating and air conditioning systems.
Requires dedicated software and a computer interface to download data to a PC. Two versions of software are available, EVSW for general use and ideal for HACCP and other quality systems, EVSWPRO which aids compliance with 21CFR Part 11.
For single channel operation select a probe with a Lumberg connector – L suffix.
For connection up to three channels use Adaptor N2000ADP/K for Type K or N2000ADP/T for Type T and select probes with a sub-miniature connector – suffix M.
- Features four channels for up to 3 external thermocouple probes or one external thermistor probe or one external thermocouple probe with thermistor at the cold junction
- Wide measurement ranges from -200°C to thermocouple limits, 400°C type T and 1372°C type K
- LCD Display for temperature readings and alarm indication
- Tough, moulded case, dust and waterproof to IP67 standards
- Large memory capacity of up to 16000 readings
- Ability to log over multiple periods
- Single, multi-function button
- LED indicators to show active logging and alarm conditions, plus audible alarm warning
- Temperature system accuracy, suitable for food industry applications
- Long battery life
- Two year warrantyKit includes:
- 1 x N2014 Data Logger
- 1 x N2000CRU Computer Interface
- 1 x EVSW EV SoftwareEverything you need to start logging!
Detail:
Sensor Type | Thermocouple type T |
Thermocouple type K | |
Thermistor | |
|
|
Measurement Range | |
Type K | -200°C to +1372°C/-328°F to +2502°F |
Type T | -200°C to +400°C/-328°F to +752°F |
Thermistor | -40°C to +70°C/-40°F to +158°F |
|
|
Scales | °C and °F |
Resolution (Thermocouple) | 0.2°C up to and including +300°C |
0.5°C above +300°C | |
|
|
Temperature accuracy | |
+20°C ambient | ±0.5°C +0.3% of reading (Logger Only) |
-20°C to +40°C ambient | ±0.5°C +0.5% of reading (Logger Only) |
-40°C to +70°C ambient | ±1.0°C +1.0% of reading (Logger Only) |
-20°C to +40°C ambient | ±0.3°C Thermistor (System Accuracy with Thermistor) |
-40°C to +70°C ambient | ±0.5°C Thermistor (System Accuracy with Thermistor) |
|
|
Memory 32K | |
1 channel | 16000 samples |
2 channels | 8000 samples |
3 channels | 5300 samples |
4 channels | 4000 samples |
|
|
Communications | Via infra red interface |
Download Time | 3 minutes for 10,000 readings (typical) |
Logging Frequency | Programmable between 2 seconds and 99 hours |
Internal Sensor Response time | T90 = 15 to 30 minutes in ambient air |
Ambient Storage | -40°C to +70°C/-40°F to +158°F |
|
|
Logging Start | Programmable date and time or manual start from instrument multi-function button |
Software | Windows™ based software for Windows XP or Vista |
|
|
Battery Type | 1 x AA size 3.6V replaceable lithium battery |
Battery Life | Up to 5 years |
|
|
Case Material | ABS |
Environmental Protection | IP67 BS EN 60529 IEC 529 |
|
|
LED Indication | |
Red | Alarm |
Green | Logger Active |
|
|
Dimensions | |
Length | 88mm/3.5ins (including fixing lug) |
Width | 80mm/3.1ins |
Depth | 35mm/1.4ins |
EVSW Software and New USB Cradle Drivers can be downloaded using these links.
- Download EVSW Software
- Download USB Cradle Drivers For Windows XP and Win 7 32 and 64-Bit
- Download Installation Guide
Diligence EV Thermocouple Data Logger with LCD Display Accessories
N2000CRU | USB Computer Interface |
A17476 | Lithium Battery |
N2000ADP/K | K Type Adapter for N2014 |
N2000ADP/T | T Type Adapter for N2014 |