CASCADA II.I SYSTEM

Output Flow Rates

 

System

RO

Type II

Type I

Cascada II.I 5

5 L/h

≥ 2 L/min*

Up to 2 L/min

Cascada II.I 10

10 L/h

≥ 2 L/min*

Up to 2 L/min

Cascada II.I 20

20 L/h

≥ 2 L/min*

Up to 2 L/min

Cascada II.I 30

30L/h

≥ 2 L/min*

Up to 2 L/min

 

 


Output Water Quality

Water Type ll

Resistivity (MΩ • cm @ 25ºC)  : > 5
TOC(ppb)                                  : < 30
Bacteria (cfu/mL)                        : <1*
Particles (>0.2 μm)                     : <1/mL*
Endotoxin (Eu/mL)                     : < 0.001 **
RNases (ng/mL)                          : < 0.01 **
DNases (pg/μL)                          : < 4 **

Water Type l

Resistivity (MΩ • cm @ 25ºC)  : 18.2
TOC(ppb)                                  : < 5 ***
Bacteria (cfu/mL)                        : <1*
Particles (>0.2 μm)                     : <1/mL*
Endotoxin (Eu/mL)                     : < 0.001 **
RNases (ng/mL)                          : < 0.01 **
DNases (pg/μL)                          : < 4 **

*    With 0.2 μm bacteria removal final filter
**  With positive charged Nylon KA1 endotoxin removal final filter
*** With feed TOC less than 2 ppm

Feed Water Requirements

Conductivity           : <1500 μs/cm @ 25 ºC
Pressure                  : 0.5 ~ 6 bar
Temperature           : 5 ~ 35 ºC
Free Chlorine          : < 3 ppm
Silt Density Index  : <12
pH                          : 4 ~ 10

Dimension (mm)

System                      : H 575 x W 366 x D 492
Reservoir                   : H1200/900/600 x W 390 x D 384
Pre-Treatment           : H 463 x W 220 x D 380
Dispensiong Station  : H 845 x W 280 x D 300

Net Weight (kg)

System                          : 23
Reservoir                      : 5(35L) ; 7(70L) ; 9 (105L)
Pre-Treatment              : 7
Dispensiong Station     : 6

Electrical Requirements

Input Voltage : 100-240 V 50-60 Hz
Power             : 200 VA Main Unit
75 VA Pretreatment