Solves the problem of acrylic plate degradation caused by the sterilization lamp

Solves the problem of acrylic plate degradation caused by the sterilization lamp

The DX type has a device that constantly circulates dry, sterile air through an acrylic housing which can be installed with a Japan Pall Corporation-made HEPA capsule filter (dust collection efficiency: 99.97%, 0.3 µm), silica gel and activated carbon.(It is possible to sterilize the HEPA capsule with ethylene oxide gas.)

One unit can be used as a sterilized box and dry box.

Uses a sterilization-resistant lamp plate.
Big opening makes it secure to place and remove microscopes and other devices.
Door can be locked so work can progress with security.

Code Model drawing
0228 U-DX type (PDF)

Specification

Outside Dimension (mm) 995W x 575D x 530H
Pass box Inner dimension(mm) 170W x 190D x 190W
Weight Approx. 18kg
air circulating device one
gas cocks for filling 2 pieces
sterilization lamp; fluorescent lamp one of each (per one unit), 10 W
outlet double-barreled built-in outlet
Fan 50/60 Hz, 5/4 W
accessories One pair of gloves
Use as a dry box
Normally, air that has been adequately dried using silica gel is circulated, but including the use of nitrogen gas is even more effective. (Always install a decompression valve on the gas tank for use.)1. Open the cock on the upper left-hand side of the main unit.2. Inject the gas to the cock on the upper right-hand side of the main unit using the diameters 6mm tube.3. This depends on the gas flow amount, gas will fill the inside in approximately three to five minutes, so close both gas stopper cocks on the tank. 

Use as a sterile box
When using this under normal conditions, the sterilizing lamp is adequate. However, dry, sterilized air can be constantly circulated inside using an air filter device (removes more than 99.97% of the particles of 0.3 µm using the DOP test method) and silica gel (filled to approximately 80% of the silica gel tank).

1. Close the main unit door, and turn on the switches for the sterilizing lamp and fan.

2. When the sterilizing lamp is on, always mount the accessory shielding plate.

3. When using the pass box on the right of the main unit, open the outside door while the inside door is closed, and open the inside door after closing the outside door.

4. If the filter or silica gel is dirty, clogged or discolored, replace them.