For routine to research analysis, this instrument series offers real innovations that provide enhanced usability and performance – without added complexity. With straightforward software, cutting-edge instrumentation and a wide selection of accessories, the Evolution 201 and 220 systems work with you to perform experiments the way you want – for the results you need.

INSIGHT Software—Sophistication Simplified

  • Personalize your home screen for easy navigation and convenience
  • Workbook tools save time and ensure accurate results
  • CUE scripting allows creation of customized interface and simplified workflow

Optimized Engineering Delivers Ultimate Performance

  • Double-beam geometry with 1nm spectral bandwidth for precision measurements
  • Application Focused Beam Geometry (AFBG) technology enables the best performance from microcells, solid samples, and fiber optic probes
  • Xenon flash lamp requires no warm-up time allowing instant measurements
  • Electronic triggering provides the highest level of accuracy for rapid-mixing kinetics measurements
  • Precision monochromator drive delivers fast scanning data collection and wavelength accuracy

Powerful Convenience at your Fingertips

  • Color, touch screen provides instrument control using built-in computer
  • Integrated keypad allows communication with INSIGHT software
  • Choice of Local or PC configurations provides flexibility and convenience

Accessories for all your Sampling Needs

  • Wide accessory selection with various temperature control options, sample changers, sippers and solid sampling devices complete your analysis
  • Choice of Peltier Single Cell Holder or a Smart 8-Cell Peltier system for temperature control and sample monitoring from 0° to 100 °C
  • Easy software integration gives you complete control of your measurements

Assurance of System Performance

  • The Thermo Scientific UV Validator package provides support for system qualification and validation activities
  • Documentation facilitates compliance of your system with the requirements of FDA,GxP, ISO 9001:2000 and ISPE 2001 guidelines and regulations
  • Mercury Lamp accessory is available to perform instrument calibration
  • Thermo Scientific INSIGHT security software provides complete tools to meet 21 CFR Part 11 requirements
Model Evolution 201 Evolution 220
Spectral Bandwidth 1nm Variable: 1nm; 2nmAFBG Microcell optimized

AFBG Fiber optic optimized

AFBG Materials optimized

Optical Design Double Beam with sample and reference cuvet positions; Czerny-Turner Monochromator
Light Source Xenon flash lamp, 3-year warranty (Typical lifetime: 5 years)
Detector Dual Silicon Photodiodes
Wavelength
Range 190 to 1100nm
Accuracy ±0.8nm (full range 190 to 1100nm)± (546.11nm mercury line)
Scan Speed <1 to 6000nm/min.; variable
Photometric
Range >3.5A
Accuracy 0.5A: ±0.0004A1A: ±0.006A

2A: ±0.010A

Measured at 440nm using neutral density filters traceable to NIST*/NLPL

Noise 0A: ≤0.00015 A1A: ≤0.00050 A

2A: ≤0.00080 A

260nm, 1nm SBW, RMS

Dimensions 24.3 x 19.1 x 11 in. (62.2 x 48.6 x 27.9mm)
Weight 31.7 lb. (14.4kg)
Electrical Requirements 90 to 240VAC 47 – 63Hz, Single Fuse (110V vs. 220V)

 

Cat. No. Description
840-210800 Evolution 201 PC
840-210700 Evolution 201 LC
840-210600 Evolution 220 PC
840-210500 Evolution 220 LC