Orion* AquaMate 7000 Vis Spectrophotometers
The new Thermo Scientific* Orion AquaMate 7000 Vis Spectrophotometer offers accuracy, dependability and ease of use.



The Orion AquaMate 7000 Vis Spectrophotometer is the ideal instrument to have in your lab for water analysis and beyond. This innovative spectrophotometer features a compact design and is intuitive to operate. When a 325 to 1100nm wavelength measurement is required, the AquaMate 7000 is exactly what you need for your water and wastewater analysis.
Thermo Scientific Orion AquaMate 7000 Vis Spectrophotometer Features:
- Includes over 260 preprogrammed methods for easy and convenient measurements using Orion AQUAfast*, Merck*, CHEMetrics* or Hach* chemistries
- Preprogrammed methods can also be adapted for new chemistries or operators can create their own custom methods
- Instruments allow a one point adjustment on any preprogrammed method to correct for variations in batch reagent chemistries
- A wide variety of vials sizes can be used, including 10mm square vials, 20 and 50mm rectangular vials and an adjustable round vial holder that accommodates 13, 16 and 24mm round vials
- Three USB connections provide the options to store methods and data to a USB memory stick, transfer data to a computer and print hard copy data to an external printer
- Compatible with ink jet and laser printers with HP PCL 5 or greater
- Includes performance verification tests that ensure wavelength accuracy and instrument functionality
- In accordance with GLP and GMP, each verification report gives the time, date and instrument serial number
- The built-in wavelength accuracy test is compatible with either the internal lamp or external calibrated standards
- To validate the instrument performance further, built-in test methods for stray light, noise and resolution are available
Includes:
- Methods on USB memory stick
- 3-position cell holder
- User documentation
- Dust cover
- USB cable
- 110V and 220V power cords
Product Specifications | |
---|---|
Optical Design | Single beam |
Spectral Bandwidth | 5.0nm |
Light Source (Typical Lifetime) | Tungsten-halogen lamp (1000 hours) |
Detector | Silicon photodiode |
Wavelength | |
Range | 325 to 1100nm |
Accuracy | ±1.0nm |
Repeatability | ±0.5nm |
Slew Speed | 11,000nm/min |
Scanning Speed | 10 to 4,200nm/min |
Data Interval | 1.0, 2.0, 3.0, 5.0nm |
Photometric | |
Measurement Modes | Absorbance, % transmittance, concentration |
Range | -0.1 to 3.0 A ; -0.3 to 125 %T ; ±9999 C |
Linearity | Up to 3.0 A at 340nm |
Accuracy | 0.5 % or ±0.005A, whichever is greater up to 2A |
Noise | <0.001 A at 0.0 A, <0.002 A at 2.0 A Peak-to peak at 340nm |
Drift | 0.002 A/hr after warm-up |
Stray Light | <0.08 %T at 220nm and 340nm, <0.1 %T at 340nm and 400nm, <1.0 %T at 198nm |
Display | Graphical with LCD backlight—9.7 x 7.1cm (3.8 x 2.8 in.) |
Keypad | Sealed membrane with tactile response keys |
Connectivity | USB type A port for USB stick (front panel), USB type B port for computer (rear panel), USB type A port for printer (rear panel) |
Dimensions | 40L x 30W x 25cm H (15.7 x 11.8 x 9.8 in.) |
Weight | 8.6kg (19 lb.) |
Power Requirements | 100 to 240V ; 50 to 60Hz |